ERI TM 600

The First Preclinical Whole-body ERI Scanner

 

ERI TM 600

The First Preclinical Whole-body ERI Scanner

 
ERI TM600 with work table Novilet
General specifications
  • Main magnet strength: 2,2mT
  • Gradient strength: 1,3 mT/cm (13 G)
  • Modulation amplitude:  5mT (50 G)
  • Resonance frequency 575 MHZ
  • Bore size: 34 mm
Dimensions and space requirements
  • Main tower: H135 x W73 x D63cm
  • Cooling tower: W110 x D84 x H180
  • Minimum space requirements: 6 m2
  • 16A type C, 3-phase power supply

Technical Specifications

Combining many of our solutions, we have made our scanner the fastest and most sensitive device on the market.

The spatial resolution of up to 600µm ensures the best quality of image captures.

The 2-seconds examination guarantees obtaining a 3D spin probe distribution map (225 projections).

The detector, which has a diameter of 34mm, is large enough to easily fit a whole mouse or small rat.

The high sensitivity of the ERI TM 600 allows the use of a low concentration of a marker, which is beneficial for animals.

Simple and user-friendly software that is suitable even for inexperienced users.

ERI TM600

Technical Specifications

Combining many of our solutions, we have made our scanner the fastest and most sensitive device on the market.

General specifications
  • Main magnet strength: 0.022T
  • Gradient strength: 13 Gs/cm
  • Modulation amplitude: 40 Gs
  • Measurement volume: 20 cm3
  • Bore size: 34 mm
Power and cooling requirements
  • 16A type C, 3-phase power supply
  • Room temperature below 24 oC
  • Air cooling capacity for chiller 2,4kW
Accessories
  • Animal handling
  • Work table
  • 24" monitor
  • Keyboard / mouse
Dimensions and space requirements
  • Main tower: H135xW73xD63cm
  • Cooling tower: H135xW52xD70cm
  • Minimum space requirements: 6 m2

Find out more abouttechnical specifications

Take up a new scientific challenge!

The spatial resolution of up to 600µm ensures the best quality of image captures.

The 2-seconds examination guarantees obtaining a 3D spin probe distribution map (225 projections).

The high sensitivity of the ERI TM 600 allows the use of a low concentration of a marker, which is beneficial for animals.

The detector, which has a diameter of 38mm, is large enough to easily fit a whole mouse or small rat.

Simple and user-friendly software that is suitable even for inexperienced users.

Take up a new scientific challenge!

Sensitivity od L-Band

Animal Bed

  • Temperature control by integrated hot-air channels avoiding the side effect of the open airflow (dehydration of the eyes, contamination by pathogens) etc.
  • Built-in integrated tubes for easy anaesthetic dose ended with a cone-shaped mouse head positioning system

Set of equipment of L-Band EPR spectroscopy novilet

Measurement bore

  • 34 mm diameter measurement bore dedicated to whole-body mouse examination 
                      
  • High-resolution automatic bed positioning system gives full control of the distance to the detection area
Image
  • Temperature control by integrated hot-air channels avoiding the side effect of the open airflow (dehydration of the eyes, contamination by pathogens) etc.
  • Built-in integrated tubes for easy anaesthetic dose ended with a cone shape mouse head positioning system
Image
  • 34 mm diameter measurement bore dedicated to whole-body mouse examination
  • High-resolution automatic bed positioning system gives full control of the distance to the detection area